INEFICACY OF PORTABLE DEVICE ACCUTREND® PLUS IN EQUINE INTERNAL MEDICINE / Ineficácia do aparelho portátil accutrend® plus na clínica médica de equinos.

2011; Volume: 27; Issue: 1 Linguagem: Inglês

10.15361/2175-0106.2011v27n1p017-021

ISSN

2175-0106

Autores

Antônio Raphael Teixeira Neto, R. G. Botelho, K. E. Sousa, Julianna Letícia Gimenes Cotrick Gomes, E. M. M. Lima,

Tópico(s)

Muscle metabolism and nutrition

Resumo

Comparou-se nesse estudo a determinacao de quatro constituintes sanguineos (glicose, lactato, triglicerides e colesterol), entre um aparelho portatil e metodos laboratoriais convencionais. Foram analisados o sangue de 20 equinos (11 machos e 9 femeas), de diversas racas com idades variando entre 8  5 anos e peso corporeo de 327,60  64,02 Kg. Apos exame clinico completo de cada animal realizou-se puncao da veia jugular externa para coleta de amostra sanguinea, de maneira que a mesma amostra, foi utilizada para determinacao no aparelho portatil e no laboratorio. Na comparacao entre as metodologias, em relacao a glicose, nao foi notada diferenca entre as medias (teste T de Student), a correlacao de Pearson revelou-se fraca e nao houve concordância segundo o metodo de Bland-Altman. Para o lactato tambem nao se observou concordância entre os dois metodos. Quanto aos triglicerides e colesterol foi possivel apenas a analise descritiva dos resultados, devido as concentracoes dos padroes fisiologicos de equinos em repouso, se apresentarem inferiores ao intervalo de leitura do aparelho portatil. Este trabalho demonstrou que o uso do referido aparelho para determinacao de glicose, lactato, triglicerides e colesterol nao e adequado na clinica medica equina. Summary Four blood components (glucose, lactate, triglycerides and cholesterol) were determined and compared using a portable device (Accutrend® Plus, ROCHE) and laboratory methods. Blood samples from 20 horses were analyzed (11 geldings and 9 mares, from various breeds with age and weight varying between 8 ± 5 years and 327.60 ± 64.02 Kg, respectively. After complete clinical examination, blood was withdrawn through venipunction of the external jugular and these samples were analyzed using a portable device and compared to laboratory results. Statistical analyses were used to compare both methodologies. The results showed that glucose data were not different with a weak Pearson's correlation and no agreement according to Bland-Altman method. The results obtained for lactate according to both methods were also not in agreement. Triglycerides and cholesterol data could not be compared due to the fact that reference values for resting horses lie below the reading range of the portable device. The present study showed that the use of Accutrend ® Plus to determine glucose, lactate, triglycerides, and cholesterol levels, is not adequate in equine internal medicine.

Referência(s)
Altmetric
PlumX