Artigo Revisado por pares

Retention of luting agents on implant abutments of different height and taper

2005; Wiley; Volume: 16; Issue: 5 Linguagem: Inglês

10.1111/j.1600-0501.2005.01159.x

ISSN

1600-0501

Autores

Mario Bresciano, Gianmario Schierano, Carlo Manzella, Antonio Screti, Cristina Bignardi, Giulio Preti,

Tópico(s)

Orthodontics and Dentofacial Orthopedics

Resumo

Abstract Objective: The purpose of this study was to evaluate the retention of four cements used to lute metal restorations onto computer designed and manufactured titanium implant abutments of different heights and convergence angles. Material and methods: Test specimens consisted of metal coping cemented on Procera ® titanium abutments of 5, 7, and 9 mm of height, and of 0°, 4°, and 8° of convergence angle. The cements used were zinc‐phosphate, zinc oxide‐eugenole, polyurethane resin with and without vaseline. The samples were tested in a universal testing machine. The mean uniaxial load at point of separation was determined. The results were statistically evaluated with ANOVA. Results: The luting agent, the height, and the taper influence the retention of metal casting on customized CAD–CAM implant abutments. Conclusion: The most retentive cement was zinc‐phosphate, followed by polyurethane, polyurethane plus vaseline, and zinc oxide‐eugenol.

Referência(s)