ResumoThis paper describes the use of two complementary techniques, LC-MS and NMR, to quantify a low level mono phosphate substituted impurity in an oligonucleotide drug substance. This impurity is the result of a sulphurisation failure, leading to the production of a sequence where a phosphorothioate linkage is replaced by a phosphate. Few quantitative methods are possible to analyse these challenging molecules especially if reversed phase ion pair chromatography, one of the most commonly used techniques for the separation of oligonucleotides, is unable to resolve the impurity in question. With the use of a standard addition method it could be demonstrated that both analytical techniques show equivalency and furthermore, the LC-MS method alone with additional validation has the potential to perform this quantitative assay with a high degree of accuracy.