MP 35 N: A corrosion resistant, high strength alloy for orthopedic surgical impants: Bio‐assay results
1975; Wiley; Volume: 9; Issue: 3 Linguagem: Inglês
10.1002/jbm.820090306
ISSN1097-4636
AutoresF Escalas, Jorge O. Galante, W. Rostoker, Philip S. Coogan,
Tópico(s)Hydrogen embrittlement and corrosion behaviors in metals
ResumoAbstract A cobalt based alloy, MP 35 N, with excellent mechanical properties has been recently introduced as a material for surgical orthopedic implants. A study was made of local and systemic host response to this material in two different mammal species. The implantation time ranged from one to 12 months. The result of this study indicated: MP 35 N produces a degree of local tissue response comparable to that of 316L stainless steel. No sytemic side effects were observed during the implantation times included in this study.