Electro-optical Wavelength Selection Enables Confocal Ratio Imaging at Low Light Levels

1995; Oxford University Press; Volume: 1; Issue: 1 Linguagem: Inglês

10.1017/s1431927695110016

ISSN

1435-8115

Autores

Roland Nitschke, Kenneth R. Spring,

Tópico(s)

Cell Image Analysis Techniques

Resumo

A confocal attachment (Odyssey) to an inverted microscope was modified to better study living cultured epithelial cells stained with fluorescent dyes. Improvements to the instrument included elimination of light leaks, improved electronic shielding, reduction of thermal effects, and use of low dark current detectors. In addition, rapid changes in illumination wavelength and power were accomplished by replacing the original mechanical filter changer by an acousto-optic tunable filter attached to the argon laser light source. The addition of a liquid crystal tunable filter to one of the two photomultiplier detectors also permitted rapid spectral scanning of the fluorescence emission. High-resolution, differential interference contrast transmitted light images were formed simultaneously by replacement of the photodiode-based transmitted light detector with a photomultiplier tube and dichroic mirror assembly. An illumination intensity of only 40 μW/cm 2 at the back focal plane of the microscope objective allowed high-quality fluorescence and transmitted light images of living cells at video rates with minimal bleaching and photodynamic damage. Both excitation ratio imaging and emission spectral scanning of living epithelial cells were accomplished. The system performance was evaluated by optical sections of fluorescent beads and thin films.