Development and validation of a method for quantitative determination of valsartan in human plasma by liquid chromatography-tandem mass spectrometry
2007; Elsevier BV; Volume: 43; Issue: 5 Linguagem: Inglês
10.1016/j.jpba.2006.12.030
ISSN1873-264X
AutoresNozomu Koseki, Hiroto Kawashita, Hisanori Hara, Miyuki Niina, Makoto Tanaka, Ryosei Kawai, Yusuke Nagae, Naoki Masuda,
Tópico(s)Pharmacological Effects and Assays
ResumoA sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the determination of valsartan in human plasma was developed and validated. A 0.5 ml aliquot was extracted using solid-phase extraction in an Empore high performance extraction disk plate, universal resin 96-well format. The estimated calibration range of the method was 2-2000 ng/ml. The method was fully validated with intra-day mean accuracy and precision of 94.8-107% and 2.19-5.40% and inter-day mean accuracy and precision of 93.5-105% and 1.87-5.67%, respectively. No significant loss of valsartan in processed samples was confirmed in processed samples for up to 24 h at 10 degrees C. Sample dilution up to 50-fold with blank human plasma provided acceptable analyses. No interference peaks or matrix effects were observed. No effect of QC sample location results was observed in a 96-well plate. This LC-MS/MS technique was found to improve quantitative determination of valsartan allowing its pharmacokinetic evaluation with clinically relevant doses.
Referência(s)