Artigo Acesso aberto Revisado por pares

Comparative Evaluation of the Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Version 2 Test Using the TaqMan 48 Analyzer and the Abbott RealTi m e HIV-1 Assay

2010; American Society for Microbiology; Volume: 49; Issue: 1 Linguagem: Inglês

10.1128/jcm.01285-10

ISSN

1098-660X

Autores

Estrelita Janse van Rensburg, Karen Tait, Alistair J. Watt, Robert Schall,

Tópico(s)

HIV/AIDS Research and Interventions

Resumo

Acceptable precision was achieved in a comparison study of the Abbott RealTime (RT) and Roche CAP/CTM-48 V2 HIV-1 assays, but viral load quantification was under- and overestimated, respectively, compared to the 2nd HIV-1 WHO International Standard. The same quantification patterns were observed for patient cohorts from Africa and the United States.