Artigo Revisado por pares

New UV spectral responsivity scale of the PTB based on a cryogenic radiometer and an argon plasma arc radiation source

2006; IOP Publishing; Volume: 43; Issue: 2 Linguagem: Inglês

10.1088/0026-1394/43/2/s15

ISSN

1681-7575

Autores

P. Meindl, Andreas E. Klinkmüller, Lutz Werner, Uwe Johannsen, K. Grützmacher,

Tópico(s)

Advanced Sensor Technologies Research

Resumo

A cryogenic-radiometer-based calibration facility utilizing monochromatized radiation of an argon arc plasma has been put into operation at the PTB. The high radiance of the plasma source allows us to improve the accuracy of the UV spectral responsivity scale between 200 nm and 410 nm and to achieve relative standard uncertainties of about 0.1% to 0.2%. Comparison with the PTB laser-based cryogenic radiometer at five laser lines in the UV indicated excellent agreement.

Referência(s)