Impedance pneumography for long‐term monitoring of respiration during sleep in adult males

1984; Wiley; Volume: 4; Issue: 4 Linguagem: Inglês

10.1111/j.1475-097x.1984.tb00808.x

ISSN

1365-2281

Autores

Vagn Larsen, Per‐Henrik Christensen, H Oxhøj, Torben Brask,

Tópico(s)

Respiratory Support and Mechanisms

Resumo

Summary. Impedance pneumography is based on the relationship between respired volume (ΔV) and changes of transthoracic impedance (Δ/Z/) during breathing. The impedance signal, the relationship between ΔV and Δ/Z/ and the success rate during long‐term recording was studied in 11 healthy adult males, aged 21–46, by a portable tape recorder (Oxford Medilog 4–24). The signals from the tape recorder (Δ/Z/), and from a spirometer (ΔV) were recorded simultaneously on a mingograph during normal breathing, apnoea, deep sighs, and coughs. The relationship between ΔV and Δ/Z/ was also analysed on an X‐Y recorder. Δ/Z/ was delayed in relation to ΔV and during one respiratory cycle the relationship was not linear. The time constant of Δ/Z/ was 6 s. A significant correlation was found between peak values of ΔV and Δ/Z/ for 0 < ΔKV<2·5 litres. Long‐term recording of Δ/Z/ during one night's sleep resulted in a success rate of 63–100%, mean 90%. Invalidation of the impedance signal was caused by overloading (81%) and interference (19%). It is concluded that the Medilog respiratory recorder enables determination of respiratory frequency, periods of apnoea and a semiquantitative estimate of volume, but artifacts may invalidate the signal.