Artigo Revisado por pares

The new cold neutron radiography and tomography instrument CONRAD at HMI Berlin

2006; Elsevier BV; Volume: 385-386; Linguagem: Inglês

10.1016/j.physb.2006.05.411

ISSN

1873-2135

Autores

André Hilger, Nikolay Kardjilov, Markus Ströbl, W. Treimer, John Banhart,

Tópico(s)

Nuclear reactor physics and engineering

Resumo

The new cold neutron radiography instrument CONRAD is a multifunctional facility for radiography and tomography with cold neutrons at Hahn–Meitner Institut, Berlin. It is located at the end of a curved neutron guide, which faces the cold-neutron source of the BER-II research reactor. The geometry provides a cold-neutron beam with wavelengths between 2 and 12 Å. Two measuring positions are available for radiography and tomography investigations. The first one is placed at the end of the guide and it is optimized for in situ experiments in which a high neutron flux is required. The available flux at this position is approximately 108 cm−2 s−1. The second measuring position uses a pin-hole geometry which allows better beam collimation (L/D up to 1000) and higher image resolution in the range of 200 μm in the CCD based detector system (10×10 cm2). The use of cold neutrons for radiography purposes increases the image contrast and improves the sensibility e.g., the detection of small amounts of water and hydrogen-containing materials in metal matrixes. On the other hand the cold-neutron beam can be modified easily by using diffraction and neutron optical techniques. This enables to perform radiography and tomography experiments with more sophisticated measuring techniques. Recent examples of research and industrial applications will be presented.

Referência(s)
Altmetric
PlumX