Construção e aplicação de biossensores usando diferentes procedimentos de imobilização da peroxidase de vegetal em matriz de quitosana
2006; Brazilian Chemical Society; Volume: 29; Issue: 5 Linguagem: Inglês
10.1590/s0100-40422006000500009
ISSN1678-7064
AutoresInês Rosane Welter Zwirtes de Oliveira, Iolanda Cruz Vieira,
Tópico(s)Electrochemical sensors and biosensors
ResumoVEGETABLE PEROXIDASE ON CHITOSAN MATRIX.Biosensors were developed by immobilization of gilo (Solanum gilo) enzymatic extract on chitosan biopolymers using three different procedures: glutaraldehyde, carbodiimide/ glutaraldehyde and epichlorohydrin/glutaraldehyde.The best biosensor performance was obtained after the immobilization of peroxidase on chitosan with epichlorohydrin/glutaraldehyde.Linear analytical curves for hydroquinone concentrations from 2.5x10 -4 to 4.5x10 -3 mol L -1 with a detection limit of 2.0x10 -6 mol L -1 and recovery of hydroquinone ranging from 95.1 to 105% were obtained.The relative standard deviation was < 1.0 % for a solution of 3.0x10 -4 mol L -1 hydroquinone and 2.0x10 -3 mol L -1 hydrogen peroxide in 0.1 mol L -1 phosphate buffer solution at pH 7.0 (n=8).The lifetime of this biosensor was 6 months (at least 300 determinations).
Referência(s)