Avaliação de ICP OES com configuração axial ou radial para determinação de iodo em sal de cozinha
2012; Brazilian Chemical Society; Volume: 35; Issue: 7 Linguagem: Inglês
10.1590/s0100-40422012000700004
ISSN1678-7064
AutoresAdriana A. Oliveira, Joaquim A. Nóbrega, Edenir Rodrigues Pereira‐Filho, Lilian C. Trevizan,
Tópico(s)Analytical Chemistry and Chromatography
ResumoTABLE SALT.The performance of inductively-coupled plasma optical emission spectrometers with axial and radial views for determination of iodine in table salt was evaluated.Interference and memory effects in nitric acid and water-soluble tertiary amines (CFA-C) media were studied.Based on a factorial experiment, one optimum instrument operational condition for axial configuration, and two optima conditions for radial configuration was established.The ICP OES with axial view was 5-fold more sensitive than the radial view.Both matrix matching and standard addition methods were used for iodine quantification and for most samples, both strategies of calibration led to similar results.Recoveries ranged from 104 to 114%.