Calculation of true coincidence summing corrections for extended sources with EFFTRAN
2011; Elsevier BV; Volume: 69; Issue: 6 Linguagem: Inglês
10.1016/j.apradiso.2011.02.042
ISSN1872-9800
AutoresT. Vidmar, Günter Kanisch, Gaj Vidmar,
Tópico(s)Radiation Detection and Scintillator Technologies
ResumoA deterministic code was developed for the calculation of true coincidence summing correction factors and has been incorporated into the EFFTRAN tool. The approach is aimed at the analysis of extended samples measured on p-type HPGe detectors in environmental gamma-ray spectrometry and was verified against the results of a state-of-the-art full Monte Carlo code. The two sets of results matched on average within 1%. Our code requires no measurements in addition to a standard full-energy-peak calibration, has a very short run time and takes into account the spatial variation of the effciency across the sample volume. The EFFTRAN code is free software, available from the authors on request.