Multiwire X-ray detector systems at the daresbury srs
1988; Elsevier BV; Volume: 273; Issue: 2-3 Linguagem: Inglês
10.1016/0168-9002(88)90094-0
ISSN1872-9576
AutoresRobert A. Lewis, Isaiah Sumner, A. Berry, J. Bordas, A. Gabriel, G. Mant, B. Parker, Kevin J. Roberts, J. S. Worgan,
Tópico(s)Particle Accelerators and Free-Electron Lasers
ResumoSynchrotron radiation allows the study of structural dynamics by X-ray diffraction techniques. These experiments impose unusual demands on X-ray detector systems. High count rates, wide dynamic range, good spatial and temporal resolution and real time data processing are all required. A recently completed system fulfills many of these requirements having a spatial resolution of better than 0.3 × 0.3 mm over an area of 200 × 200 mm with a 600 kHz throughput and a dynamic range limited by counting statistics. Examples derived from the existing experimental programme are used to illustrate the performance of these devices.
Referência(s)