Multicenter validation of the magnetic resonance t2* technique for segmental and global quantification of myocardial iron
2009; Wiley; Volume: 30; Issue: 1 Linguagem: Inglês
10.1002/jmri.21781
ISSN1522-2586
AutoresAnna Ramazzotti, Alessia Pepe, Vincenzo Positano, Giuseppe Rossi, Daniele De Marchi, Maria Gabriella Brizi, A. Luciani, Massimo Midiri, Giuseppina Sallustio, Gianluca Valeri, Vincenzo Caruso, Michele Centra, Paolo Cianciulli, Vincenzo De Sanctis, Aurelio Maggio, Massimo Lombardi,
Tópico(s)Bone and Joint Diseases
ResumoAbstract Purpose To assess the transferability of the magnetic resonance imaging (MRI) multislice multiecho T2* technique for global and segmental measurement of iron overload in thalassemia patients. Materials and Methods Multiecho T2* sequences were installed on six MRI scanners. Five healthy subjects ( n = 30) were scanned at each site; five thalassemia major (TM) patients were scanned at the reference site and were rescanned locally ( n = 25) within 1 month. T2* images were analyzed using previously validated software. Results T2* values of healthy subjects showed intersite homogeneity. On TM patients, for global heart T2* values the correlation coefficient was 0.97, coefficients of variation (CoV s ) ranged from 0.04–0.12, and intraclass coefficients (ICC s ) ranged from 0.94–0.99. The mean CoV and ICC for segmental T2* distribution were 0.198 and 88, respectively. Conclusion The multislice multiecho T2* technique is transferable among scanners with good reproducibility. J. Magn. Reson. Imaging 2009;30:62–68. © 2009 Wiley‐Liss, Inc.
Referência(s)