Artigo Revisado por pares

Oxygen transfer performance of a new Sarns membrane oxygenator

1992; SAGE Publishing; Volume: 7; Issue: 4 Linguagem: Inglês

10.1177/026765919200700404

ISSN

1477-111X

Autores

David W Fried, John Bell-Thomson,

Tópico(s)

Cardiac Structural Anomalies and Repair

Resumo

Oxygen transfer performance of the SMO 2 , a new hollow-fibre membrane oxygenator from Sarns/3M, was evaluated. The maximum oxygen transfer of the SMO 2 (504 ml O 2 /min) was 57% higher than that for the SMO. This was true even though the membrane surface area increased by only 11 % (1.8 m2 to 2.0 m2). The average gas phase-arterial oxygen gradient and the average shunt fraction was decreased by 35% and 30% respectively. The diffusing capacity of the SMO 2 was increased by 74% over the SMO. The oxygen transfer performance compared favourably, not only to the SMO, but to five other previously evaluated membrane oxygenators. It performed better in each category regardless of whether the result was expressed in absolute terms or relative to membrane surface area.

Referência(s)
Altmetric
PlumX