Effects of scintillator on the detective quantum efficiency (DQE) of a digital imaging system
2005; Elsevier BV; Volume: 101; Issue: 2 Linguagem: Inglês
10.1016/j.tripleo.2005.07.032
ISSN1528-395X
AutoresTaeko T. Farman, Robert H. Vandre, John C. Pajak, Stuart Miller, Alex Lempicki, Allan G. Farman,
Tópico(s)Medical Imaging Techniques and Applications
ResumoTo compare the effects of scintillator on the detective quantum efficiency (DQE) of a charge-coupled device (CCD) digital intraoral radiographic system.Three screens composed of 3 different scintillator materials, namely europium-doped lutetium oxide (Lu2(O3):Eu3+), transparent optical ceramic (TOC), thallium-doped cesium iodide (CsI:Tl; CsI), and terbium-doped gadolinium oxysulfide (Gd2(O2)S:Tb; GOS) were compared, in turn, in combination with a CCD detector having square pixels with height and width dimensions of 19.5 microm. DQE was investigated using the slanted-slit-derived MTF and surrogate signal-to-noise ratio (SNR) measurements derived from calculations of the mean and standard deviations from the mean pixel values of multiple random patches from various uniform exposures. An Irix x-ray generator operated at 70 kVp and 8 mA, with a nominal focal spot size of 0.7 mm and 2.5 mm Al equivalent filtration, was used in making all exposures.Using TOC, the peak DQE was 62% at 5 cycles/mm. For CsI, the peak DQE was 22% at 2 cycles/mm. With GOS, the peak DQE was 10% at 1 cycle/mm.Under identical experimental settings, TOC consistently resulted in higher DQE than CsI and commercially available GOS scintillators combined with the same high-resolution solid-state detector.
Referência(s)