Pupillenperimetrie mit dem OCTOPUS 1-2-3 - Erste Erfahrungen
1994; Thieme Medical Publishers (Germany); Volume: 204; Issue: 05 Linguagem: Inglês
10.1055/s-2008-1035567
ISSN1439-3999
AutoresStephanie Turtschi, Oliver Bergamin, Barbara Dubler, Andreas Schötzau, Mario Zulauf,
Tópico(s)Corneal surgery and disorders
ResumoDevelopment of a clinically useful pupil perimetry on an OCTOPUS 1-2-3.Various test parameters were investigated on a moderately modified OCTOPUS 1-2-3. We evaluated the pupillary responses of 122 eyes of 122 healthy volunteers qualitatively.Pleasing pupillary responses were obtained with the following parameters: 200 ms stimuli, Goldmann size 5 (1.5 degrees), background illumination 3 apostilb. Acoustic signals 0.5 seconds before stimulus presentation were helpful to minimize blinking artifacts. The pupillary reactions were fare stronger in central areas of the visual field than in mid-peripheral areas (i.e. 26 degrees).Pupil perimetry can be preformed on a moderately modified OCTOPUS 1-2-3. The results reveal considerable inter- and intra-individual variability. Equal pupillary responses at all eccentricities would improve subsequent mathematical evaluation.
Referência(s)