Artigo Revisado por pares

Autoactivation of source dwell positions for HDR brachytherapy treatment planning

2000; Wiley; Volume: 27; Issue: 11 Linguagem: Inglês

10.1118/1.1315315

ISSN

2473-4209

Autores

Stavroula Giannouli, Dimos Baltas, Natasa Milickovic, Michael Lahanas, Christos Kolotas, Nikolaos Zamboglou, Nikolaos Uzunoglu,

Tópico(s)

Photoacoustic and Ultrasonic Imaging

Resumo

The most accurate classical dose optimization algorithms in HDR brachytherapy strongly depend on an appropriate selection of source dwell positions which fulfill user‐defined geometrical boundary conditions which are relative to patient anatomy. Most anatomical situations, such as for prostate and head and neck tumors, are complex and can require geometries with 5–15 catheters with 48 possible dwell positions per catheter depending on the tumor volume. The manual selection of dwell positions using visual checks by trial and error is very time consuming. This can only be improved by the use of a technique which automatically recognizes and selects the optimum dwell positions for each catheter. We have developed an algorithm, termed an autoactivation algorithm, which improves implant planning by providing a facility for the necessary automatic recognition of HDR source dwell positions.

Referência(s)
Altmetric
PlumX