A Determination of the Volt
1989; IOP Publishing; Volume: 26; Issue: 1 Linguagem: Inglês
10.1088/0026-1394/26/1/003
ISSN1681-7575
AutoresW K Clothier, G. J. Sloggett, H Bairnsfather, Malcolm F. Currey, D.J. Benjamin,
Tópico(s)Radioactive Decay and Measurement Techniques
ResumoThe volt has been determined to high accuracy using a liquid electrometer. The representation of the volt maintained by the Josephson effect using the value of 2 e/h = 483.594 THz/V recommended by the Consultative Committee on Electricity of the International Committee of Weights and Measures in 1972 has been measured to be smaller than the SI volt by (8.09 ± 0.27) ppm. This implies that the SI value of 2 e/h is (483.59791 ± 0.00013) THz/V. Full experimental details are given.
Referência(s)