Artigo Revisado por pares

End-compensated magnetostatic cavity for polarized H3e neutron spin filters

2009; American Institute of Physics; Volume: 80; Issue: 6 Linguagem: Inglês

10.1063/1.3154386

ISSN

1527-2400

Autores

James McIver, R. W. Erwin, Wangchun Chen, T. Gentile,

Tópico(s)

Advanced NMR Techniques and Applications

Resumo

We have expanded upon the “Magic Box” concept, a coil driven magnetic parallel plate capacitor constructed out of mu-metal, by introducing compensation sections at the ends of the box that are tuned to limit end-effects similar to those of short solenoids. This ability has reduced the length of the magic box design without sacrificing any loss in field homogeneity, making the device far more applicable to the often space limited neutron beam line. The appeal of the design beyond affording longer polarized H3e lifetimes is that it provides a vertical guide field, which facilitates neutron spin transport for typical polarized beam experiments. We have constructed two end-compensated magic boxes of dimensions 28.4×40×15 cm3 (length×width×height) with measured, normalized volume-averaged transverse field gradients ranging from 3.3×10−4 to 6.3×10−4 cm−1 for cell sizes ranging from 8.1×6.0 to 12.0×7.9 cm2 (diameter×length), respectively.

Referência(s)