Artigo Revisado por pares

Improved Arrhythmia Detection in Implantable Loop Recorders

2008; Wiley; Volume: 19; Issue: 9 Linguagem: Inglês

10.1111/j.1540-8167.2008.01156.x

ISSN

1540-8167

Autores

Michele Brignole, Carissa L. Bellardine Black, Poul Erik Bloch Thomsen, Richard Sutton, Andrés Moyá, Robert Stadler, Jian Cao, Marc Messier, Heikki V. Huikuri,

Tópico(s)

ECG Monitoring and Analysis

Resumo

Implantable loop recorders (ILR) have an automatic arrhythmia detection feature that can be compromised by inappropriately detected episodes. This study evaluated a new ILR sensing and detection scheme for automatically detecting asystole, bradyarrhythmia, and tachyarrhythmia events, which is implemented in the next generation device (Reveal DX/XT).The new scheme employs an automatically adjusting R-wave sensing threshold, enhanced noise rejection, and algorithms to detect asystole, bradyarrhythmia, and tachyarrhythmia. Performance of the new algorithms was evaluated using 2,613 previously recorded, automatically detected Reveal Plus episodes from 533 patients. A total of 71.9% of episodes were inappropriately detected by the original ILR, and at least 88.6% of patients had one or more inappropriate episodes, with most inappropriate detections due to R-wave amplitude reductions, amplifier saturation, and T-wave oversensing. With the new scheme, inappropriate detections were reduced by 85.2% (P < 0.001), with a small reduction in the detection of appropriate episodes (1.7%, P < 0.001). The new scheme avoided inappropriate detections in 67.4% of patients that had them with the original scheme.The new sensing and detection scheme is expected to substantially reduce the occurrence of inappropriately detected episodes, relative to that of the original ILR.

Referência(s)