Precision performance at low levels and 99th percentile concentration of the Access® AccuTnI® assay on two different platforms
2009; De Gruyter; Volume: 47; Issue: 3 Linguagem: Inglês
10.1515/cclm.2009.080
ISSN1437-4331
AutoresMartina Zaninotto, Monica Mion, E Novello, Marco Moretti, Ernesto Delprete, Marco Rocchi, Davide Sisti, Mario Plebani,
Tópico(s)Cardiac electrophysiology and arrhythmias
ResumoAbstract : Cardiac troponins currently represent the preferred biomarkers for the detection of myocardial necrosis. The objective of the present study was to compare the performance of the Access : The Access : cTnI concentrations are given in μg/L. 10% CV values (95% confidence interval, CI) were 0.0577 (0.0467–0.0750) (UniCel : Our data confirm the reliability of the evaluated cTnI assay and demonstrate the comparability of the cTnI values between the platforms studied. Clin Chem Lab Med 2009;47:367–71.
Referência(s)