Artigo Revisado por pares

Study of RPC gas composition using Daya Bay RPCs

2010; IOP Publishing; Volume: 34; Issue: 8 Linguagem: Inglês

10.1088/1674-1137/34/8/015

ISSN

2058-6132

Autores

Liehua Ma, Y. F. Wang, Zhang Jiawen, L. Lebanowski, Viktor Pěč, Xu Guang-Hua, Jin Chen, Y. G. Xie, Jifeng Han, Zhang Qing‐min, S. Qian, Zhe Ning,

Tópico(s)

Radiation Detection and Scintillator Technologies

Resumo

This study uses Daya Bay RPCs operating in streamer mode to investigate gas mixtures of at least 50% argon, at most 6% isobutane, and with small amounts of SF6. Isobutane is reduced to 2% without degradation of performance, and SF6 reduces the noise rate and current, as well as the signal size. This study provides quantitative relationships between basic RPC operating parameters and various gas compositions.