Point-of-Care Vertical Flow Allergen Microarray Assay: Proof of Concept

2014; American Association for Clinical Chemistry; Volume: 60; Issue: 9 Linguagem: Inglês

10.1373/clinchem.2014.223230

ISSN

1530-8561

Autores

Thiruppathiraja Chinnasamy, Loes I. Segerink, Mats Nystrand, Jesper Gantelius, Helene Andersson Svahn,

Tópico(s)

Advanced biosensing and bioanalysis techniques

Resumo

Abstract BACKGROUND Sophisticated equipment, lengthy protocols, and skilled operators are required to perform protein microarray-based affinity assays. Consequently, novel tools are needed to bring biomarkers and biomarker panels into clinical use in different settings. Here, we describe a novel paper-based vertical flow microarray (VFM) system with a multiplexing capacity of at least 1480 microspot binding sites, colorimetric readout, high sensitivity, and assay time of <10 min before imaging and data analysis. METHOD Affinity binders were deposited on nitrocellulose membranes by conventional microarray printing. Buffers and reagents were applied vertically by use of a flow controlled syringe pump. As a clinical model system, we analyzed 31 precharacterized human serum samples using the array system with 10 allergen components to detect specific IgE reactivities. We detected bound analytes using gold nanoparticle conjugates with assay time of ≤10 min. Microarray images were captured by a consumer-grade flatbed scanner. RESULTS A sensitivity of 1 ng/mL was demonstrated with the VFM assay with colorimetric readout. The reproducibility (CV) of the system was <14%. The observed concordance with a clinical assay, ImmunoCAP, was R2 = 0.89 (n = 31). CONCLUSIONS In this proof-of-concept study, we demonstrated that the VFM assay, which combines features from protein microarrays and paper-based colorimetric systems, could offer an interesting alternative for future highly multiplexed affinity point-of-care testing.

Referência(s)