Deuterium oxide as a reagent for the modification of mass spectra in electrospray microcolumn liquid chromatography—mass spectrometry

1993; Elsevier BV; Volume: 647; Issue: 1 Linguagem: Inglês

10.1016/0021-9673(93)83321-i

ISSN

1873-3778

Autores

K. Karlsson,

Tópico(s)

Microfluidic and Capillary Electrophoresis Applications

Resumo

Microcolumn liquid chromatography was used in combination with a quadrupole mass spectrometer equipped with an electrospray ionization interface. Deuterium oxide was used as a reagent to induce peak shifts in mass spectra. The peak shift obtained gives the number of heteroatoms carrying hydrogen atoms that are present in a molecule. Examples relating to amino alcohols, peptides and sugars are given. Two different modes of operation were investigated; postcolumn addition of the reagent to the eluent from the column and the use of deuterium oxide as a mobile phase component. The latter technique makes use of a dynamic exchange reaction on-column. Exchanged hydrogen atoms are carried away from the migrating zone (as H2HO or H2O) containing the target compound, leading to a high yield of the deuterated compound. On-column exchange was found to be more useful, especially if the compounds contained several exchangeable hydrogen atoms. No significant changes were observed with regard to chromatographic selectivity or efficiency.