A fully automated amino acid analyzer using NBD‐F as a fluorescent derivatization reagent
2004; Wiley; Volume: 18; Issue: 9 Linguagem: Inglês
10.1002/bmc.365
ISSN1099-0801
AutoresChiaki Aoyama, Tomofumi Santa, Makoto Tsunoda, Takeshi Fukushima, Chieko Kitada, Kazuhiro Imai,
Tópico(s)Mass Spectrometry Techniques and Applications
ResumoA fully automated amino acid analyzer using NBD-F (4- fluoro-7-nitro-2,1,3-benzoxadiazole) as a fluorescent derivatization reagent was developed. The whole analytical process was fully automated from derivatization, injection to HPLC separation and quantitation. The derivatization reaction conditions were re-evaluated and optimized. Amino acids were derivatized by NBD-F for 40 min at room temperature in the borate buffer (pH 9.5). The derivatives were separated within 100 min and fluorometrically detected at 540 nm with excitation at 470 nm. The detection limits for amino acids were in the range of 2.8-20 fmol. The calibration curves were linear over the range of 20 fmol to 20 pmol on column with the correlation coefficients of 0.999. The coefficients of variation were less than 5% at 3 pmol injection for all amino acids. Amino acids in rat plasma were determined by the proposed HPLC method.