Construct Validity Testing of a Laparoscopic Surgical Simulator

2006; Lippincott Williams & Wilkins; Volume: 202; Issue: 5 Linguagem: Inglês

10.1016/j.jamcollsurg.2006.01.004

ISSN

1879-1190

Autores

Elspeth M. McDougall, Federico A. Corica, John Boker, L Sala, Gabriela Stoliar, James F. Borin, Frank T. Chu, Ralph V. Clayman,

Tópico(s)

Augmented Reality Applications

Resumo

We present initial data on the construct, content, and face validity of the LAPMentor (Simbionix), virtual reality laparoscopic surgical simulator.Medical students (MS), residents and fellows (R/F), and experienced laparoscopic surgeons (ES), with < 30 laparoscopic cases per year (ES 30 laparoscopic cases per year (ES>30), were tested on 9 basic skill tasks (SK) including manipulation of 0-degree and 30-degree cameras (SK1, SK2), eye-hand coordination (SK3), clipping (SK4), grasping and clipping (SK5), two-handed maneuvers (SK6), cutting (SK7), fulguration (SK8), and object-translocation (SK9).Mean MS (n=23), R/F (n=24), ES 30 (n=30) ages were 26 years (range 21 to 32 years), 31 years (range 27 to 39 years), 49 years (range 31 to 70 years) and 47 years (range 34 to 69 years), respectively. In the lower level skill tasks (SK3, SK4, SK5, and SK6) the ES>30, ES 30 scores tended to be better than the R/F and ES 30 had notably higher SK8 scores than the R/F and ES<30, who had similar scores, and these had notably better scores than the MS.The noncamera skills (SK3 to 9) of the LAPMentor surgical simulator can distinguish between laparoscopically naive and ES. SK8 showed the highest level of construct validity, by accurately differentiating among the MS, R/F, ES 30.