Artigo Produção Nacional Revisado por pares

Cyclic Fatigue of Protaper Instruments

2006; Elsevier BV; Volume: 33; Issue: 1 Linguagem: Inglês

10.1016/j.joen.2006.09.003

ISSN

1878-3554

Autores

Hélio Pereira Lopes, Edson Jorge Lima Moreira, Carlos Nelson Elias, Renata Andriola de Almeida, Mônica Schultz Neves,

Tópico(s)

Dental materials and restorations

Resumo

The present work evaluated the influence of the curved segment length of artificial root canals (the arc) and the number of cycles necessary to fracture engine-driven nickel-titanium endodontic instruments. ProTaper F3 25-mm files at 250 rpm were used in two artificial canals. The artificial canals were made of stainless steel with an inner diameter of 1.04 mm, a total length of 20 mm, and arc on the ends with a radius of curvature of 6 mm. The arc length of the first tube measured 9.4 mm, and the straight part measured 10.6 mm. The second tube was 14.1 mm long, and the straight part measured 5.9 mm. We determined the fracture surface distances and the number of cycles necessary to induce fatigue fracture in the ProTaper F3 instruments. The fracture surfaces and the helical shaft of the instruments were investigated using a scanning electron microscope. The results indicated that the required number of cycles to cause a fracture was influenced by the canal arc length, the morphology of the fractured surface presented ductile characteristics, and plastic deformation in the helical shaft of the fractured instruments did not occur.