Artigo Revisado por pares

Modification of a whole-body NMR imager into a radio frequency EPR spectrometer suitable for in vivo measurements

1996; IOP Publishing; Volume: 7; Issue: 7 Linguagem: Inglês

10.1088/0957-0233/7/7/003

ISSN

1361-6501

Autores

Stephen McCallum, Marcello Alecci, David J. Lurie,

Tópico(s)

Advanced NMR Techniques and Applications

Resumo

We report the modification of a low-field whole-body NMR imager to allow radio frequency EPR spectroscopy. The instrument is designed to give optimum sensitivity for in vivo detection of free radicals. The RF circuit is able to operate over a wide frequency range (240 - 320 MHz) and is designed to handle input power levels of up to 12.5 W. The EPR resonator is of the loop - gap type suitable for samples up to 400 ml. A remote resonator coupling method has been developed enabling convenient matching adjustment. An automatic frequency control circuit is able to adjust for frequency deviations caused by animal motion. Where possible, existing imager hardware and commercially available instruments have been used. The instrument is controlled from a central computer via an IEEE 488 instrumentation bus. Here we present sensitivity measurements obtained from a variety of large aqueous samples containing nitroxide free radicals. We show that the instrument is suitable for the detection of exogenous free radicals in 200 g rats.

Referência(s)