High power ultrasound standard

2002; Acoustical Society of America; Volume: 111; Issue: 4 Linguagem: Inglês

10.1121/1.1459464

ISSN

1520-9024

Autores

George S. K. Wong, Lixue Wu,

Tópico(s)

Ultrasound Imaging and Elastography

Resumo

A sensitive radiation force balance for laboratory measurement of ultrasonic power is presented. The principle of the system is based on measuring the ultrasonic radiation force exerted on a conical float suspended in water. Technical details of the implementation of the economically attractive system are described. The operation of the system is automated with the aid of the IEEE-488 bus and a desktop computer. Design aspects that affect measurement uncertainty are investigated. A theoretical model for the measurement of ultrasonic power with a conical reflector target is discussed. The expanded uncertainty (95% confidence level) of the above radiation force conical float system is estimated to be between 5% to 10%.

Referência(s)