Artigo Acesso aberto Revisado por pares

Heterophilic Antibodies Remain a Problem for the Immunoassay Laboratory

1997; Oxford University Press; Volume: 108; Issue: 4 Linguagem: Inglês

10.1093/ajcp/108.4.417

ISSN

1943-7722

Autores

Greg Ward, Letitia Mckinnon, Tony Badrick, Peter E. Hickman,

Tópico(s)

Blood disorders and treatments

Resumo

The prevalence of heterophilic antibody interference in a modern immunochemiluminometric assay containing blocking agents was determined using thyrotropin as an illustrative example. Serum samples were obtained from 295 consecutive patients who underwent routine thyroid function testing. The following versions of the thyrotropin assay were used: protocol A (zero blocker), protocol B (routine blocker concentration), and protocol C (extra blocker). Ten patients (prevalence 3.4%) had significant levels of heterophilic antibodies (protocol A value greater than 9 SD from the protocol B value). The observed thyrotropin levels for protocols B and C were the same for all patients, consistent with the reagent blockers in routine assays adequately eliminating heterophilic antibody interference. However, seven more patients (0.03%) in series of 21,000 assessed by routine thyroid function testing had discordant results because of a concentration of heterophilic antibodies so high as to overwhelm the added blocking agents.

Referência(s)