The Fricke dosimeter as a 1-hit detector
1986; Pergamon Press; Volume: 11; Issue: 6 Linguagem: Inglês
10.1016/1359-0189(86)90056-7
ISSN1878-1691
AutoresRobert S. Katz, G.L. Sinclair, M.P.R. Waligórski,
Tópico(s)Radiation Therapy and Dosimetry
ResumoA fit to the experimental data for the response of the Fricke dosimeter to energetic heavy ions is obtained using a calculation of the relative effectiveness of a 1-hit detector, from track theory. We use 2 fitted parameters, the target size, α0, which may be thought to represent a ‘diffusion length’, and E0, the dose of gamma-rays at which there is an average of one hit per target (the D-37 dose), and a new algorithm for the average radial distribution of dose in liquid water from the passing ion. The G value for ions is then given as the product of the calculated relative effectiveness and the experimental G value for gamma rays.