The First Trial of Phase Contrast Imaging for Digital Full-Field Mammography Using a Practical Molybdenum X-Ray Tube

2005; Lippincott Williams & Wilkins; Volume: 40; Issue: 7 Linguagem: Inglês

10.1097/01.rli.0000165575.43381.48

ISSN

1536-0210

Autores

Toyohiko Tanaka, Chika Honda, Satoru Matsuo, Kazuo Noma, Hiromu Oohara, Norihisa Nitta, Shinichi Ota, Keiko Tsuchiya, Yoko Sakashita, Aya Yamada, Michio Yamasaki, Akira Furukawa, Masashi Takahashi, Kiyoshi Murata,

Tópico(s)

Advanced X-ray and CT Imaging

Resumo

The image quality of a newly developed full-field digital phase contrast mammography (PCM) system and of a conventional screen-film (SF) mammography system were compared via images of a phantom and receiver operating characteristic (ROC) analysis of clinical images.Magnified (1.75X) PCM images were scanned (sampling rate, 43.75 microm) and then reduced to original-sized, 25-micron pixel images printed on photothermographic film. Along with corresponding SF images, the phantom images were evaluated subjectively, and the clinical images of 38 patients were subjected to ROC analysis of mass and microcalcification.In the image quality of a phantom, the PCM exceeded the SF. In both mass and microcalcification, the ROC analysis Az values of the PCM clinical images surpassed those of the SF images.The PCM provides better images than the SF. Clinical trials suggest superior detection of both mass and microcalcification by full-field digital PCM over conventional SF mammography.