Artigo Revisado por pares

Dosimetric characteristics of a 6 MV photon beam from a linear accelerator with asymmetric collimator jaws

1988; Elsevier BV; Volume: 14; Issue: 2 Linguagem: Inglês

10.1016/0360-3016(88)90447-6

ISSN

1879-355X

Autores

Jatinder Palta, Komanduri M. Ayyangar, N. Suntharalingam,

Tópico(s)

Radiation Dose and Imaging

Resumo

Dosimetric measurements have been made of a 6 MV photon beam from a linear accelerator equipped with asymmetric jaws. The field size factors for asymmetrically set fields are compared to those for symmetrically set fields. The change of beam quality has been measured as a function of off axis position of the asymmetric fields to assess its effect on depth dose. Additional measurements include beam penumbra and shape of isodose curves for open and wedge fields as the field opening is moved asymmetrically from the central ray.

Referência(s)