Consequences of misfilling contemporary vaporizers with desflurane
1993; Springer Science+Business Media; Volume: 40; Issue: 1 Linguagem: Inglês
10.1007/bf03009323
ISSN1496-8975
AutoresJeff J. Andrews, Robert Johnston, George C. Kramer,
Tópico(s)Cardiac, Anesthesia and Surgical Outcomes
ResumoDesflurane is a volatile anaesthetic that combines low blood gas solubility (blood/gas partition coefficient =0.42 at 37° C), moderate potency (MAC = 6–7%), and high volatility (vapour pressure =681 mmHg at 20° C, boiling point = 23.5° C). The volatility and potency of desflurane prevent its safe use in vaporizers of traditional design. We present a mathematical model which demonstrates the potential for desflurane overdose if contemporary vaporizers are misfilled with desflurane. The most hazardous filling error occurs if an enflurane vaporizer is misfilled with desflurane. The calculated desflurane output of a misfilled enflurane vaporizer at a dial setting of 1% and a temperature of 22° C is 57.8%, or 9.6 MAC. For misfilled enflurane, isoflurane, and halothane vaporizers at dial settings equivalent to one MAC at 22° C, the calculated desflurane output is 14.0, 10.2, and 7.8 MAC, respectively. We conclude that the safe delivery of desflurane will require engineering safeguards, additional monitoring, and education of the anesthesia community.