Artigo Revisado por pares

Automated sub-ppm mass accuracy on an ESI-TOF for use with drug discovery compound libraries

2007; American Chemical Society; Volume: 18; Issue: 9 Linguagem: Inglês

10.1016/j.jasms.2007.06.001

ISSN

1879-1123

Autores

Justin G. Stroh, Christopher Petucci, Scott J. Brecker, Nelson Huang, James M. Lau,

Tópico(s)

Pesticide Residue Analysis and Safety

Resumo

An automated, routine method to obtain sub-ppm accurate mass data on a benchtop electrospray ionization time-of-flight (ESI-TOF) mass spectrometer is described. Standards in the mass range 114 to 734 Da were analyzed over a 5-day period to demonstrate intra- and interday precision and mean mass accuracy less than 1 ppm. One hundred drug discovery pharmaceutical compounds were used to demonstrate an absolute average mass accuracy of 0.47 +/- 0.31 ppm. This is in contrast to previous reports of accurate mass analysis using time-of-flight mass spectrometry (TOFMS) technology that operates within 3 to 5 ppm. The same 100 samples were also analyzed using Fourier transform mass spectrometry (FTMS) technology and yielded comparable results to the TOFMS analysis.

Referência(s)