LC method for the determination of R-timolol in S-timolol maleate: Validation of its ability to quantify and uncertainty assessment
2005; Elsevier BV; Volume: 68; Issue: 4 Linguagem: Inglês
10.1016/j.talanta.2005.07.026
ISSN1873-3573
AutoresR.D. Marini, Patrice Chiap, Bruno Boulanger, Serge Rudaz, Eric Rozet, Jacques Crommen, Philippe Hubert,
Tópico(s)Analytical Chemistry and Chromatography
ResumoThis article presents the validation results of a chiral liquid chromatographic (LC) method previously developed for the quantitative determination of R-timolol in S-timolol maleate samples. A novel validation strategy based on the accuracy profiles was used to select the most appropriate regression model, to assess the method accuracy within well defined acceptance limits and to determine the limits of quantitation as well as the concentration range. The validation phase was completed by the investigation of the risk profiles of various acceptable regression models in order to ensure the risk of obtaining the future measurements outside the acceptance limits fixed a priori. On the other hand, the present paper also shows how data used in this validation approach can be used to estimate the measurement uncertainty. The uncertainty derived from β-expectation tolerance interval (σTol2), which is equal to the uncertainty of measurements as well as the expanded uncertainty (Ux) using a coverage factor k = 2 was estimated. The uncertainty estimates obtained from validation data were finally compared with those obtained from interlaboratory and robustness studies.
Referência(s)