Determination of Plasma Norepinephrine and Epinephrine by High Performance Liquid Chromatography Using a Two-Column System and an Electrochemical Detector
1982; Marcel Dekker; Volume: 5; Issue: 10 Linguagem: Inglês
10.1080/01483918208062865
ISSN2331-0413
AutoresG. Lachacarontre, G Nicot, C. Magnet, J. L. Rocca, L. Merle, Jean-Paul Valette, P. Pignolet,
Tópico(s)Analytical Methods in Pharmaceuticals
ResumoAbstract A method for the simultaneous plasma norepinephrine (NE) and epinephrine (E) determination by reversed-phase ion-pair liquid chromatography with electrochemical detection has been developed. Catecholamines were extracted from a 4 ml plasma sample using an alumina adsorption procedure. A two-pump, two-injection valve, two-column system allowed both to detect plasma NE and E with a good sensitivity due to large injected volumes of extract without any electrochemical detector disturbance and to eliminate uric acid and dopa the low k' of which would prevent the NE detection. Using this method, NE and E would be detected in respective injected amounts down to 30 and 50 picograms. Plasma NE and E determinations were found to be linear in the range of 288 to 788 pg/ml and 24 to 274 pg/ml respectively. The reproducibility, expressed as the coefficients of variation, varied from 2.1% for NE to 10.8% for E.
Referência(s)