Determination of local anaesthetics and their impurities in pharmaceutical preparations using HPLC method with amperometric detection
2005; Elsevier BV; Volume: 37; Issue: 5 Linguagem: Inglês
10.1016/j.jpba.2004.09.025
ISSN1873-264X
AutoresZbigniew Fijałek, Emil Baczyński, Agata Piwońska, Małgorzata Warowna-Grześkiewicz,
Tópico(s)Microfluidic and Capillary Electrophoresis Applications
ResumoA method for the determination of local anaesthetics and their impurities--2,6-dimethylaniline and o-toluidine--by high-performance liquid chromatographic method with amperometric detection has been developed. The analysis was performed in an isocratic mode on a reversed phase Luna column 5 microm C-18 (100 mm x 4.6 mm). A mobile phase [0.01 mol l(-1) Tris buffer of pH 7.9:acetonitrile (45:55)] was selected for the separation and determination of studied anaesthetics and their impurities. Chromatograms were recorded for 500 s by means of an amperometric detector at a potential of +1.0 V of the glassy carbon electrode versus the reference electrode Ag/AgCl. The proposed liquid chromatographic method was successfully applied to the analysis of commercially available pharmaceutical preparations. The limit of the detection for 2,6-dimethylaniline and o-toluidine was 0.8 ng ml(-1). The limit of quantitation, considering a signal to noise ratio was 1.5 ng ml(-1). The method developed in this study is sensitive and selective and can be applied to routine studies of pharmaceuticals in the form of cream and injection.
Referência(s)