Artigo Revisado por pares

Determination of picogram amounts of 99tc by neutron activation analysis

1972; Elsevier BV; Volume: 60; Issue: 2 Linguagem: Inglês

10.1016/s0003-2670(01)95001-9

ISSN

1873-4324

Autores

S.C. Foti, E. Delucchi, V. Akamian,

Tópico(s)

Radioactive Decay and Measurement Techniques

Resumo

A post-irradiation procedure for the determination of picogram amounts of 99Tc in triply distilled water is described. The 99Tc is irradiated in a thermal neutron flux of about 5.1013 n cm-2 sec-l to produce 16-sec 100Tc. The chemical purification of the 100Tc from other radionuclides requires 40–45 sec and the technetium yield is about 53%. The purified 100Tc is counted in a low-background β-counter. On décrit une méthode de dosage de 99Tc dans l'eau, en quantité de l'ordre du picogramme. Le procédé consiste à transformer 99Tc en 100Tc de 16 sec, à l'aide d'un flux de neutrons thermiques de 5.lO13 n cm-2 sec-1. La purification chimique de 100Tc d'avec d'autres éléments activés dure 40–45sec; le rendement en technétium est d'environ 53%. Es wird eine Methode zur Bestimmung von Pikogramm-Mengen 99Tc in Wasser beschrieben. Das Verfahren beruht auf der Umwandlung des 99Tc zu 16 sec100Tc durch Bestrahlung mit einem thermischen Neutronenfluss von 5.10s13n cm-2 sec-1. Die chemische Abtrennung des 100Tc von anderen aktivierten Elementen dauert 40–45sec, und die Ausbeute an Technetium beträgt ca.53%. Die Aktivität des gereinigten 100Tc wird mit einem Antikoinzidenz-β-Zähler bestimmt.