Multiple-slice imaging of a tissue-equivalent phantom by use of time-resolved optical tomography
2000; Optica Publishing Group; Volume: 39; Issue: 19 Linguagem: Inglês
10.1364/ao.39.003380
ISSN0003-6935
AutoresFlorian E. W. Schmidt, Jeremy C. Hebden, Elizabeth M. C. Hillman, Martin Fry, Martin Schweiger, Hamid Dehghani, David T. Delpy, Simon Arridge,
Tópico(s)Advanced MRI Techniques and Applications
ResumoFollowing several years of development the construction of a multichannel time-resolved imaging device for medical optical tomography has been completed. Images are reconstructed from time-resolved measurements by use of a scheme that employs a finite-element diffusion-based forward model and an iterative reconstruction solver. Prior to testing on clinical subjects the fully automated instrument and the reconstruction software are evaluated with tissue-equivalent phantoms. We describe our first attempt to generate multiple-slice images of a phantom without uniform properties along the axial direction, while still using a computationally fast two-dimensional reconstruction algorithm. The image quality is improved by the employment of an approximate correction method that uses scaling factors derived from the ratios of finite-element forward simulations in two and three spatial dimensions. The 32-channel system was employed to generate maps of the internal scattering and the absorption properties at 14 different transverse planes across the phantom. The images clearly reveal the locations of small inhomogeneous regions embedded within the phantom. These results were obtained by use of purely temporal data and without resource to reference measurements.
Referência(s)