Artigo Revisado por pares

Flow-injection Fourier transform infrared spectrometric analysis

1992; Elsevier BV; Volume: 261; Issue: 1-2 Linguagem: Inglês

10.1016/0003-2670(92)80175-7

ISSN

1873-4324

Autores

Miguel de la Guárdia, Salvador Garrigues, Máximo Gallignani, J.L. Burguera, M. Burguera,

Tópico(s)

Advanced Chemical Sensor Technologies

Resumo

A method has been developed for obtaining and storing the complete Fourier transform spectrum, of a sample in a flowing stream in only 19 s. A resolution of 4 cm−1 is recommended. The determination of o-xylene in xylol was employed as a test system, using hexane as a carrier and diluent in all the experiments. A classical liquid flow cell and a demountable flow-through cell were compared. The best limit of detection obtained was 0.01% (v/v). The effects of flow-rate and injection volume were studied for the two cells used. The relative standard deviation (0.4% o-xylene, n = 5) was ca. 1%