A rapid semi-automatic method for the determination of the total halide concentration in sea water by means of potentiometric titration
1970; Elsevier BV; Volume: 52; Issue: 3 Linguagem: Francês
10.1016/s0003-2670(01)82061-4
ISSN1873-4324
Autores Tópico(s)Advanced Chemical Sensor Technologies
ResumoA semi-automatic potentiometric titration of the total halide concentration in sea water with silver nitrate is described. The titration is followed with a silver electrode and the titration data, which are recorded directly on punched tape and type-writer, are evaluated by means of a computer program based on the GRAN extrapolation method. The determinations have a precision of ±0.02 % and, since many samples can be titrated simultaneously, the time for a single determination including evaluation and editing of titration data can be reduced to less than 5 min. On décrit une détermination de la concentration totale en halogénure d'une eau de mer, à l'aide de nitrate d'argent, par titrage potentiométrique semi-automatique. On utilise une électrode d'argent; les résultats sont enregistrés directement sur machine à écrire et interprétés à l'aide d'un programme d'ordinateur. La précision est de ±0.02 % et la durée totale d'un dosage peut être inférieure à 5 min, plusieurs échantillons pouvant être titrés simultanément. Es wird eine halbautomatische potentiometrische Titration der Gesamt-Halogenkonzentration in Meerwasser mit Silbernitrat beschrieben. Die Titration wird mit einer Silberelektrode verfolgt. Die Messergebnisse werden unmittelbar auf Lochstreifen und Schreibmaschine übertragen und mittels eines Rechenprogramms, das auf der Extrapolationsmethode von Gran beruht, ausgewertet. Die Bestimmungen haben eine Genauigkeit von ±0.02%. Da gleichzeitig viele Proben titriert werden können, kann die Zeit für eine einzelne Bestimmung einschliesslich Auswertung und Ausgabe der Titrationsergebnisse auf weniger als 5 min verringert werden.
Referência(s)