Artigo Acesso aberto Revisado por pares

Industrialized Functional Test for Insulin Micropumps

2011; Elsevier BV; Volume: 25; Linguagem: Inglês

10.1016/j.proeng.2011.12.195

ISSN

1877-7058

Autores

E. Chappel, R. Allendes, Fernando D. Bianchi, Giacomo Calcaterra, F. Cannehan, C. Conan, J. Lefrique, G.-L. Lettieri, S. Mefti, A. Noth, S. Proennecke, Frederic Neftel,

Tópico(s)

Pancreatic function and diabetes

Resumo

We report on original industrialized functional testing methods for insulin micropumps.A specific Design-For-Test associated with the analysis of the integrated pressure sensor signal, during a specific actuation profile, allows a complete characterization of the pump functionality in few seconds.All critical failures due to fabrication and assembly create clear and specific deviations from nominal signal.The pressure sensor sensitivity coupled to the high compression ratio and the small dead volume of the micropump lead to a leak detection sensitivity of only few nl/h for water or insulin at 37°C.

Referência(s)