High-performance liquid chromatography of the aromatase inhibitor, letrozole, and its metabolite in biological fluids with automated liquid-solid extraction and fluorescence detection
1996; Elsevier BV; Volume: 683; Issue: 2 Linguagem: Inglês
10.1016/0378-4347(96)00118-1
ISSN1872-812X
AutoresF. Marfil, V. Pineau, A. Sioufi, J. Godbillon,
Tópico(s)Effects and risks of endocrine disrupting chemicals
ResumoAn analytical method for the determination of letrozole (CGS 20 267) in plasma and of letrozole and its metabolite, CGP 44 645, in urine is described. Automated liquid-solid extraction of compounds from plasma and urine was performed on disposable 100-mg C8 columns using the ASPEC system. The separation was achieved on an ODS Hypersil C18 column using acetonitrile-phosphate buffer, pH 7, as the mobile phase at a flow-rate of 1.5 ml/min. A fluorescence detector was used for the quantitation. The excitation and emission wavelengths were 230 and 295 nm, respectively. The limits of quantitation (LOQ) of letrozole in plasma and in urine were 1.40 nmol/l (0.4 ng/ml) and 2.80 nmol/l, respectively. The respective mean recoveries and coefficient of variation (C.V.) were 96.5% (9.8%) in plasma and 104% (7.7%) in urine. The LOQ of CGP 44 645 in urine was 8.54 nmol/l (2 ng/ml). The mean recovery was 108% (6.3%). The compounds were well separated from co-extracted endogenous components and no interferences were observed at the retention times of compounds. The sensitivity of this method for letrozole in plasma should be sufficient for kinetic studies in humans with single doses of 0.5 mg and possibly less.
Referência(s)