Reliability and Validity of Videokeratoscopic Measurements

1991; Lippincott Williams & Wilkins; Volume: 68; Issue: 12 Linguagem: Inglês

10.1097/00006324-199112000-00006

ISSN

1538-9235

Autores

Gordon G. Heath, Daniel R. Gerstman, William Henry Wheeler, P. SARITA SONI, Douglas G. Horner,

Tópico(s)

Glaucoma and retinal disorders

Resumo

When a test surface or a cornea is carefully and properly positioned with respect to the Corneal Modeling System (CMS) videokeratoscope, measurements are highly repeatable, but the accuracy of the measurements is dependent on the shape of the surface. For spherical and toroidal surfaces, the CMS provides suitably accurate descriptions of the topography of the surface. For aspheric surfaces the CMS renders less accurate, but still clinically acceptable, topographic descriptions.