Purification of liquid xenon and impurity monitoring for a PET detector

1994; Elsevier BV; Volume: 349; Issue: 2-3 Linguagem: Inglês

10.1016/0168-9002(94)91217-3

ISSN

1872-9576

Autores

V. Chepel, I. Lopes, R. Ferreira‐Marques, A. Policarpo,

Tópico(s)

Dark Matter and Cosmic Phenomena

Resumo

The purification system and the method of impurity monitoring developed for a positron emission tomography (PET) scanner based on liquid xenon position sensitive detectors are described. The present results show that an attenuation length of the scintillation photons in the liquid of the order of 50 cm can be reached by using only commercial purifiers. Although our preliminary measurements show that a strong electron attachment exists in liquid xenon, the electron attenuation length of 2 cm achieved is sufficient for the PET detectors.