A validation study comparing accelerator MS and liquid scintillation counting for analysis of 14C-labelled drugs in plasma, urine and faecal extracts
2000; Elsevier BV; Volume: 24; Issue: 2 Linguagem: Inglês
10.1016/s0731-7085(00)00397-6
ISSN1873-264X
AutoresR. Colin Garner, James Barker, C Flavell, John Garner, M Whattam, Graeme Young, N. J. CUSSANS, S. G. Jezequel, David Tai Leong,
Tópico(s)Analytical Methods in Pharmaceuticals
ResumoA comparison has been made between accelerator mass spectrometry (AMS) analysis and liquid scintillation counting (LSC) of plasma, urine and faecal samples containing 14C-labelled drugs. In an in vitro study in which human plasma was spiked (the term spiked is used in Section 2.6) with 14C-Fluconazole (14C-FL) over a concentration range of 0.1-2.5 dpm/ml, a correlation coefficient of 0.999 was determined for AMS analysis versus extrapolated LSC data. No significant day to day (or inter-day)variation was seen (P < 0.05 by ANOVA). Coefficients of variation for these analyses ranged from 2.68 to 6.50%. In vivo studies in which rats were given a high (11.5 microCi/kg) or low (18.1 nCi/kg) radioactive dose (to model an exposure of 0.9 microSievert to man) of 14C-Fluticasone propionate(14C-FP) showed that there was also a good correspondence between AMS and LSC data. A mass balance study in a single the faeces by 96 h; less than 1% of the administered dose was excreted in the urine. The limit of reliable measurement of drug related material, above background concentrations, by AMS analysis in this study was approximately 0.1 dpm/ml for plasma, 0.01 dpm/ml for urine without any sample extraction or concentration and 0.01 dpm/ml for faecal extracts. The data reported here demonstrate that AMS is an ultrasensitive and reliable method for analysing 14C-labelled drugs in human and animal body fluids.
Referência(s)