Performance of the New Single-Bore High-Pressure Pneumatic Nebulizer (SBHPPN) in Plasma Atomic Emission Spectrometry
1994; SAGE Publishing; Volume: 48; Issue: 5 Linguagem: Inglês
10.1366/0003702944924772
ISSN1943-3530
AutoresJosé Luis Todolí Torró, Marı́a Muñoz, Manuel Valiente, Vicente Hernandis, Antonio Canals,
Tópico(s)Diamond and Carbon-based Materials Research
ResumoThe analytical performance of the new single-bore high-pressure pneumatic nebulizer (SBHPPN) has been evaluated in ICP-AES. It has been compared with a commercially available pneumatic all-glass concentric nebulizer. The new SBHPPN generates finer and narrower primary aerosols than the concentric nebulizer. With respect to the concentric nebulizer, the analyte transport rate for SBHPPN is 1.8 to 3.4 times higher, its sensitivity 2.8 to 3.8 times higher, and its limits of detection 2 to 5 times lower. The better performance shown by the SBHPPN has been verified in both the continuous and the discrete mode, as well as with high-salt-content solutions, showing no tendency to clog because of salt deposition.